5-137
IM 710105-17E
Commands
5
:MEASure:{CHANnel<x>|MATH<x>}:AREA2:
DELay:STATe
Function Sets or queries the on/off status and the display
format of a delay parameter in Area2.
Syntax
:MEASure:{CHANnel<x>|MATH<x>}:AREA2:D
ELay:STATe {OFF|ON|DEGRee}
:MEASure:{CHANnel<x>|MATH<x>}:AREA2:D
ELay:STATe?
<x> of CHANnel<x> = 1 to 4 (1 or 2 for 2-channel
models)
<x> of MATH<x> = 1 or 2 (1 only for 2-channel
models)
Example
:MEASURE:CHANNEL1:AREA2:DELAY:STA
TE ON
:MEASURE:CHANNEL1:AREA2:DELAY:STATE?
-> :MEASURE:CHANNEL1:AREA2:DELAY:STA
TE ON
:MEASure:{CHANnel<x>|MATH<x>}:<Param
eter>?
Function Queries the setting of a waveform parameter of a
waveform.
Syntax
:MEASure:{CHANnel<x>|MATH<x>}:<Parame
ter>?
<x> of CHANnel<x> = 1 to 4 (1 or 2 for 2-channel
models)
<x> of MATH<x> = 1 or 2 (1 only for 2-channel
models)
<Parameter> = {AMPLitude|
AVERage|AVGFre
q|AVGPeriod|BWIDth|DELay|DT|DUTYcycl
e|ENUMber|FALL|FREQuency|HIGH|LOW|MA
Ximum|MINimum|NOVershoot|NWIDth|PERi
od|PNUMber|POVershoot|PTOPeak|PWIDth|
RISE|RMS|SDEViation|TY1Integ|TY2Integ
|V1|V2
}
:MEASure:{CHANnel<x>|MATH<x>}:<Param
eter>:COUNt?
Function Queries the statistical processing count of a
waveform parameter.
Syntax
:MEASure:{CHANnel<x>|MATH<x>}:<Parame
ter>:COUNt?
<x> of CHANnel<x> = 1 to 4 (1 or 2 for 2-channel
models)
<x> of MATH<x> = 1 or 2 (1 only for 2-channel
models)
<Parameter> = {AMPLitude|
AVERage|AVGFre
q|AVGPeriod|BWIDth|DELay|DT|DUTYcycl
e|ENUMber|FALL|FREQuency|HIGH|LOW|MA
Ximum|MINimum|NOVershoot|NWIDth|PERi
od|PNUMber|POVershoot|PTOPeak|PWIDth|
RISE|RMS|SDEViation|TY1Integ|TY2Integ
|V1|V2
}
Example (Below is an example for the average of CH1.)
:MEASURE:CHANNEL1:AVERAGE:COUNT?
-> :MEASURE:CHANNEL1:AVERAGE:COUNT 10
0
:MEASure:{CHANnel<x>|MATH<x>}:<Param
eter>:{MAXimum|MEAN|MINimum|SDEViati
on}?
Function Queries a statistical value of a waveform
parameter.
Syntax
:MEASure:{CHANnel<x>|MATH<x>}:<Param
eter>:{MAXimum|MEAN|MINimum|SDEViati
on}?
<x> of CHANnel<x> = 1 to 4 (1 or 2 for 2-channel
models)
<x> of MATH<x> = 1 or 2 (1 only for 2-channel
models)
<Parameter> = {AMPLitude|
AVERage|AVGFre
q|AVGPeriod|BWIDth|DELay|DT|DUTYcycl
e|ENUMber|FALL|FREQuency|HIGH|LOW|MA
Ximum|MINimum|NOVershoot|NWIDth|PERi
od|PNUMber|POVershoot|PTOPeak|PWIDth|
RISE|RMS|SDEViation|TY1Integ|TY2Integ
|V1|V2
}
Example (Below is an example for the average of CH1.)
:MEASURE:CHANNEL1:AVERAGE:MAXIMUM?
-> :MEASURE:CHANNEL1:AVERAGE:
MAXIMUM 10.0000E+03
Description If the statistical value is immeasurable, the
DLM2000 returns “NAN” (not a number).
5.20 MEASure Group