5-200
IM 710105-17E
:SERialbus<x>:SENT:SEARch[:SETup]:FD
ATa:DATA<y>:CONDition
Function Sets or queries the comparison condition of the
specified user data of the SENT signal search fast
channel.
Syntax
:SERialbus<x>:SENT:SEARch[:SETup]:FDA
Ta:DATA<y>:CONDition {BETWeen|EQUal|G
REater|LESS|NOTBetween|NOTEqul}
:SERialbus<x>:SENT:SEARch[:SETup]:FDA
Ta:DATA<y>:CONDition?
<x> = 1 to 4
<y> = 1 to 4
Example
:SERIALBUS1:SENT:SEARCH:SETUP:FDATA:D
ATA1:CONDITION BETWEEN
:SERIALBUS1:SENT:SEARCH:SETUP:FDATA:D
ATA1:CONDITION?
-> :SERIALBUS1:SENT:SEARCH:SETUP:FDAT
A:DATA1:CONDITION BETWEEN
:SERialbus<x>:SENT:SEARch[:SETup]:FD
ATa:DATA<y>:DECimal<z>
Function Sets or queries the data of the specified user data
of the SENT signal search fast channel in decimal
notation.
Syntax
:SERialbus<x>:SENT:SEARch[:SETup]:FDA
Ta:DATA<y>:DECimal<z> {<NRf>}
:SERialbus<x>:SENT:SEARch[:SETup]:FDA
Ta:DATA<y>:DECimal<z>?
<x> = 1 to 4
<y> = 1 to 4
<z> = 1 or 2
<NRf> = See the DLM2000 Features Guide for
this information.
Example
:SERIALBUS1:SENT:SEARCH:SETUP:FDATA:D
ATA1:DECIMAL1 1
:SERIALBUS1:SENT:SEARCH:SETUP:FDATA:D
ATA1:DECIMAL1?
-> :SERIALBUS1:SENT:SEARCH:SETUP:FDAT
A:DATA1:DECIMAL1 1
:SERialbus<x>:SENT:SEARch[:SETup]:FD
ATa:DATA<y>:MODE
Function Sets or queries the data enable/disable condition
of the specified user data of the SENT signal
search fast channel.
Syntax
:SERialbus<x>:SENT:SEARch[:SETup]:FDA
Ta:DATA<y>:MODE {<Boolean>}
:SERialbus<x>:SENT:SEARch[:SETup]:FDA
Ta:DATA<y>:MODE?
<x> = 1 to 4
<y> = 1 to 4
Example
:SERIALBUS1:SENT:SEARCH:SETUP:FDATA:D
ATA1:MODE ON
:SERIALBUS1:SENT:SEARCH:SETUP:FDATA:D
ATA1:MODE?
-> :SERIALBUS1:SENT:SEARCH:SETUP:FDAT
A:DATA1:MODE 1
:SERialbus<x>:SENT:SEARch[:SETup]:FD
ATa:DNIBbles?
Function Queries all fast channel nibble data settings for
SENT signal searching.
Syntax
:SERialbus<x>:SENT:SEARch:[SETup]:FDA
Ta:DNIBbles?
<x> = 1 to 4
:SERialbus<x>:SENT:SEARch[:SETup]:FD
ATa:DNIBbles:CONDition
Function Sets or queries the nibble data comparison
condition of the SENT signal search fast channel.
Syntax
:SERialbus<x>:SENT:SEARch[:SETup]:FDA
Ta:DNIBbles:CONDition {FALSe|TRUE}
:SERialbus<x>:SENT:SEARch[:SETup]:FDA
Ta:DNIBbles:CONDition?
<x> = 1 to 4
Example
:SERIALBUS1:SENT:SEARCH:SETUP:FDATA:D
NIBBLES:CONDITION TRUE
:SERIALBUS1:SENT:SEARCH:SETUP:FDATA:D
NIBBLES:CONDITION?
-> :SERIALBUS1:SENT:SEARCH:SETUP:FDAT
A:DNIBBLES:CONDITION TRUE
:SERialbus<x>:SENT:SEARch:[SETup]:FD
ATa:DNIBbles:HEXa
Function Sets the SENT signal search fast channel nibble
data in hexadecimal notation.
Syntax
:SERialbus<x>:SENT:SEARch:[SETup]:FDA
Ta:DNIBbles:HEXa {<String>}
<x> = 1 to 4
Example
:SERIALBUS1:SENT:SEARCH:SETUP:FDATA:D
NIBBLES:HEXA "112233"
:SERialbus<x>:SENT:SEARch:[SETup]:FD
ATa:DNIBbles:PATTern
Function Sets or queries the SENT signal search fast
channel nibble data in binary notation.
Syntax
:SERialbus<x>:SENT:SEARch:[SETup]:FDA
Ta:DNIBbles:PATTern {<String>}
:SERialbus<x>:SENT:SEARch[:SETup]:FDA
Ta:DNIBbles:PATTern?
<x> = 1 to 4
Example
:SERIALBUS1:SENT:SEARCH:S
ETUP:FDATA:DNIBBLES:PATTE
RN "111100001111000011110000"
:SERIALBUS1:SENT:SEARCH:SETUP:FDATA:D
NIBBLES:PATTERN?
-> :SERIALBUS1:SENT:SEARCH:SETUP:
FDATA:DNIBBLES:PATTERN "1111000011110
00011110000"
5.24 SERialbus Group