5-287
IM 710105-17E
Commands
1
2
3
4
5
6
App
Index
:XY<x>:MEASure:CURSor:X<y>:POSition
Function Sets or queries a horizontal cursor position in the
XY display.
Syntax
:XY<x>:MEASure:CURSor:X<y>:POSiti
on {<NRf>}
:XY<x>:MEASure:CURSor:X<y>:POSition?
<x> = 1 or 2 (1 only for 2-channel models)
<y> = 1 or 2
<NRf>=−4to4(divisions)
Example
:XY1:MEASURE:CURSOR:X1:POSITION 1
:XY1:MEASURE:CURSOR:X1:POSITION?
-> :XY1:MEASURE:CURSOR:X1:POSITION 1.
000E+00
:XY<x>:MEASure:CURSor:X<y>:STATe
Function Sets or queries the on/off status of the voltage of
an X-Y display horizontal cursor.
Syntax
:XY<x>:MEASure:CURSor:X<y>:STA
Te {<Boolean>}
:XY<x>:MEASure:CURSor:X<y>:STATe?
<x> = 1 or 2 (1 only for 2-channel models)
<y> = 1 or 2
Example
:XY1:MEASURE:CURSOR:X1:STATE ON
:XY1:MEASURE:CURSOR:X1:STATE?
-> :XY1:MEASURE:CURSOR:X1:STATE 1
:XY<x>:MEASure:CURSor:X<y>:VALue?
Function Queries the voltage of a horizontal cursor in the
XY display.
Syntax
:XY<x>:MEASure:CURSor:X<y>:VALue?
<x> = 1 or 2 (1 only for 2-channel models)
<y> = 1 or 2
Example
:XY1:MEASURE:CURSOR:X1:VALUE?
-> :XY1:MEASURE:CURSOR:X1:VALUE 1.000
E+00
:XY<x>:MEASure:CURSor:Y<y>?
Function Queries all vertical cursor settings for the XY
display.
Syntax
:XY<x>:MEASure:CURSor:Y<y>?
<x> = 1 or 2 (1 only for 2-channel models)
<y> = 1 or 2
:XY<x>:MEASure:CURSor:Y<y>:POSition
Function Sets or queries a vertical cursor position in the XY
display.
Syntax
:XY<x>:MEASure:CURSor:Y<y>:POSiti
on {<NRf>}
:XY<x>:MEASure:CURSor:Y<y>:POSition?
<x> = 1 or 2 (1 only for 2-channel models)
<y> = 1 or 2
<NRf>=−4to4(divisions)
Example
:XY1:MEASURE:CURSOR:Y1:POSITION 1
:XY1:MEASURE:CURSOR:Y1:POSITION?
-> :XY1:MEASURE:CURSOR:Y1:POSITION 1.
000E+00
:XY<x>:MEASure:CURSor:Y<y>:STATe
Function Sets or queries the on/off status of the voltage of
an X-Y display vertical cursor.
Syntax
:XY<x>:MEASure:CURSor:Y<y>:STA
Te {<Boolean>}
:XY<x>:MEASure:CURSor:Y<y>:STATe?
<x> = 1 or 2 (1 only for 2-channel models)
<y> = 1 or 2
Example
:XY1:MEASURE:CURSOR:Y1:STATE ON
:XY1:MEASURE:CURSOR:Y1:STATE?
-> :XY1:MEASURE:CURSOR:Y1:STATE 1
:XY<x>:MEASure:CURSor:Y<y>:VALue?
Function Queries the voltage of a vertical cursor in the XY
display.
Syntax
:XY<x>:MEASure:CURSor:Y<y>:VALue?
<x> = 1 or 2 (1 only for 2-channel models)
<y> = 1 or 2
Example
:XY1:MEASURE:CURSOR:Y1:VALUE?
-> :XY1:MEASURE:CURSOR:Y1:VALUE 1.000
E+00
:XY<x>:MEASure:INTeg?
Function Queries all integration settings for the XY display.
Syntax
:XY<x>:MEASure:INTeg?
<x> = 1 or 2 (1 only for 2-channel models)
:XY<x>:MEASure:INTeg:LOOP
Function Sets or queries the integration method for the XY
display.
Syntax
:XY<x>:MEASure:INTeg:LOOP {CLOSe|OPEN
}
:XY<x>:MEASure:INTeg:LOOP?
<x> = 1 or 2 (1 only for 2-channel models)
Example
:XY1:MEASURE:INTEG:LOOP CLOSE
:XY1:MEASURE:INTEG:LOOP?
-> :XY1:MEASURE:INTEG:LOOP CLOSE
5.36 XY Group