EasyManuals Logo

YOKOGAWA DLM2000 Series User Manual

YOKOGAWA DLM2000 Series
340 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #320 background imageLoading...
Page #320 background image
5-287
IM 710105-17E
Commands
1
2
3
4
5
6
App
Index
:XY<x>:MEASure:CURSor:X<y>:POSition
Function Sets or queries a horizontal cursor position in the
XY display.
Syntax
:XY<x>:MEASure:CURSor:X<y>:POSiti
on {<NRf>}
:XY<x>:MEASure:CURSor:X<y>:POSition?
<x> = 1 or 2 (1 only for 2-channel models)
<y> = 1 or 2
<NRf>=−4to4(divisions)
Example
:XY1:MEASURE:CURSOR:X1:POSITION 1
:XY1:MEASURE:CURSOR:X1:POSITION?
-> :XY1:MEASURE:CURSOR:X1:POSITION 1.
000E+00
:XY<x>:MEASure:CURSor:X<y>:STATe
Function Sets or queries the on/off status of the voltage of
an X-Y display horizontal cursor.
Syntax
:XY<x>:MEASure:CURSor:X<y>:STA
Te {<Boolean>}
:XY<x>:MEASure:CURSor:X<y>:STATe?
<x> = 1 or 2 (1 only for 2-channel models)
<y> = 1 or 2
Example
:XY1:MEASURE:CURSOR:X1:STATE ON
:XY1:MEASURE:CURSOR:X1:STATE?
-> :XY1:MEASURE:CURSOR:X1:STATE 1
:XY<x>:MEASure:CURSor:X<y>:VALue?
Function Queries the voltage of a horizontal cursor in the
XY display.
Syntax
:XY<x>:MEASure:CURSor:X<y>:VALue?
<x> = 1 or 2 (1 only for 2-channel models)
<y> = 1 or 2
Example
:XY1:MEASURE:CURSOR:X1:VALUE?
-> :XY1:MEASURE:CURSOR:X1:VALUE 1.000
E+00
:XY<x>:MEASure:CURSor:Y<y>?
Function Queries all vertical cursor settings for the XY
display.
Syntax
:XY<x>:MEASure:CURSor:Y<y>?
<x> = 1 or 2 (1 only for 2-channel models)
<y> = 1 or 2
:XY<x>:MEASure:CURSor:Y<y>:POSition
Function Sets or queries a vertical cursor position in the XY
display.
Syntax
:XY<x>:MEASure:CURSor:Y<y>:POSiti
on {<NRf>}
:XY<x>:MEASure:CURSor:Y<y>:POSition?
<x> = 1 or 2 (1 only for 2-channel models)
<y> = 1 or 2
<NRf>=−4to4(divisions)
Example
:XY1:MEASURE:CURSOR:Y1:POSITION 1
:XY1:MEASURE:CURSOR:Y1:POSITION?
-> :XY1:MEASURE:CURSOR:Y1:POSITION 1.
000E+00
:XY<x>:MEASure:CURSor:Y<y>:STATe
Function Sets or queries the on/off status of the voltage of
an X-Y display vertical cursor.
Syntax
:XY<x>:MEASure:CURSor:Y<y>:STA
Te {<Boolean>}
:XY<x>:MEASure:CURSor:Y<y>:STATe?
<x> = 1 or 2 (1 only for 2-channel models)
<y> = 1 or 2
Example
:XY1:MEASURE:CURSOR:Y1:STATE ON
:XY1:MEASURE:CURSOR:Y1:STATE?
-> :XY1:MEASURE:CURSOR:Y1:STATE 1
:XY<x>:MEASure:CURSor:Y<y>:VALue?
Function Queries the voltage of a vertical cursor in the XY
display.
Syntax
:XY<x>:MEASure:CURSor:Y<y>:VALue?
<x> = 1 or 2 (1 only for 2-channel models)
<y> = 1 or 2
Example
:XY1:MEASURE:CURSOR:Y1:VALUE?
-> :XY1:MEASURE:CURSOR:Y1:VALUE 1.000
E+00
:XY<x>:MEASure:INTeg?
Function Queries all integration settings for the XY display.
Syntax
:XY<x>:MEASure:INTeg?
<x> = 1 or 2 (1 only for 2-channel models)
:XY<x>:MEASure:INTeg:LOOP
Function Sets or queries the integration method for the XY
display.
Syntax
:XY<x>:MEASure:INTeg:LOOP {CLOSe|OPEN
}
:XY<x>:MEASure:INTeg:LOOP?
<x> = 1 or 2 (1 only for 2-channel models)
Example
:XY1:MEASURE:INTEG:LOOP CLOSE
:XY1:MEASURE:INTEG:LOOP?
-> :XY1:MEASURE:INTEG:LOOP CLOSE
5.36 XY Group

Table of Contents

Other manuals for YOKOGAWA DLM2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DLM2000 Series and is the answer not in the manual?

YOKOGAWA DLM2000 Series Specifications

General IconGeneral
BrandYOKOGAWA
ModelDLM2000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals