5-60
IM 710105-17E
:ANALysis:PANalyze<x>:I2T:SCALe:CENT
er
Function Sets or queries the center value for manual
scaling.
Syntax
:ANALysis:PANalyze<x>:I2T:SCALe:CENT
er {<NRf>}
:ANALysis:PANalyze<x>:I2T:SCALe:CENT
er?
<x> = 1 or 2
<NRf>=−1.0000E+31to1.0000E+31
Example
:ANALYSIS:PANALYZE1:I2T:SCALE:CENT
ER 1
:ANALYSIS:PANALYZE1:I2T:SCALE:CENTER?
-> :ANALYSIS:PANALYZE1:I2T:SCALE:CENT
ER 1.00000E+00
:ANALysis:PANalyze<x>:I2T:SCALe:MODE
Function Sets or queries the scaling mode.
Syntax
:ANALysis:PANalyze<x>:I2T:SCALe:MO
DE {AUTO|MANual}
:ANALysis:PANalyze<x>:I2T:SCALe:MODE?
<x> = 1 or 2
Example
:ANALYSIS:PANALYZE1:I2T:SCALE:MO
DE AUTO
:ANALYSIS:PANALYZE1:I2T:SCALE:MODE?
-> :ANALYSIS:PANALYZE1:I2T:SCALE:MO
DE AUTO
:ANALysis:PANalyze<x>:I2T:SCALe:SENS
itivity
Function Sets or queries the span from the center value for
manual scaling.
Syntax
:ANALysis:PANalyze<x>:I2T:SCALe:SENSi
tivity {<NRf>}
:ANALysis:PANalyze<x>:I2T:SCALe:SENSi
tivity?
<x> = 1 or 2
<NRf>=−1.0000E+31to1.0000E+31
Example
:ANALYSIS:PANALYZE1:I2T:SCALE:SENSITI
VITY 10
:ANALYSIS:PANALYZE1:I2T:SCALE:SENSITI
VITY?
-> :ANALYSIS:PANALYZE1:I2T:SCALE:SENS
ITIVITY 10.0000E+00
:ANALysis:PANalyze<x>:I2T:TRANge
(Time Range)
Function Sets or queries the measurement time period.
Syntax
:ANALysis:PANalyze<x>:I2T:TRAN
ge {<NRf>,<NRf>}
:ANALysis:PANalyze<x>:I2T:TRANge?
<x> = 1 or 2
<NRf>=−5to5divisions(instepsof10divisions/
display record length)
Example
:ANALYSIS:PANALYZE1:I2T:TRANGE−4,4
:ANALYSIS:PANALYZE1:I2T:TRANGE?
->:ANALYSIS:PANALYZE1:I2T:TRANGE−4.
00E+00,4.00E+00
:ANALysis:PANalyze<x>:HARMonics?
Function Queries all harmonic analysis settings.
Syntax
:ANALysis:PANalyze<x>:HARMonics?
<x> = 1 or 2
:ANALysis:PANalyze<x>:HARMonics:CCLa
ss?
Function Queries all class C harmonic analysis settings.
Syntax
:ANALysis:PANalyze<x>:HARMonics:CCLa
ss?
<x> = 1 or 2
:ANALysis:PANalyze<x>:HARMonics:CCLa
ss:GETLambda
Function Queries the current power factor for class C
harmonic analysis.
Syntax
:ANALysis:PANalyze<x>:HARMonics:CCLas
s:GETLambda
<x> = 1 or 2
Example
:ANALYSIS:PANALYZE1:HARMONICS:CCLASS:
GETLAMBDA
:ANALysis:PANalyze<x>:HARMonics:CCLa
ss:LAMBda
Function Sets or queries the power factor for class C
harmonic analysis.
Syntax
:ANALysis:PANalyze<x>:HARMonics:CCLas
s:LAMBda {<NRf>}
:ANALysis:PANalyze<x>:HARMonics:CCLas
s:LAMBda?
<x> = 1 or 2
<NRf> = 0.001 to 1
Example
:ANALYSIS:PANALYZE1:HARMONICS:CCLASS:
LAMBDA 0.10
:ANALYSIS:PANALYZE1:HARMONICS:CCLASS:
LAMBDA?
-> :ANALYSIS:PANALYZE1:HARMONICS:CCLA
SS:LAMBDA100.0E−03
5.3 ANALysis Group

Table of Contents

Other manuals for YOKOGAWA DLM2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DLM2000 Series and is the answer not in the manual?

YOKOGAWA DLM2000 Series Specifications

General IconGeneral
BrandYOKOGAWA
ModelDLM2000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals