5-59
IM 710105-17E
Commands
5
:ANALysis:PANalyze<x>:I2T:MEASure:I2
T:{MAXimum|MEAN|MINimum|SDEViation}?
Function Queries a statistical value of a Joule integral.
Syntax
:ANALysis:PANalyze<x>:I2T:MEASure:I2T
:{MAXimum|MEAN|MINimum|
SDEViation}?
<x> = 1 or 2
Example
:ANALYSIS:PANALYZE1:I2T:MEASURE:I2T:M
AXIMUM?
-> :ANALYSIS:PANALYZE1:I2T:MEASURE:I2
T:MAXIMUM 10.0000E+03
Description If the statistical value is immeasurable, the
DLM2000 returns “NAN” (not a number).
:ANALysis:PANalyze<x>:I2T:MEASure:I2
T:STATe
Function Sets or queries whether or not a Joule integral will
be measured.
Syntax
:ANALysis:PANalyze<x>:I2T:MEASure:I2T
:STATe {<Boolean>}
:ANALysis:PANalyze<x>:I2T:MEASure:I2T
:STATe?
<x> = 1 or 2
Example
:ANALYSIS:PANALYZE1:I2T:MEASURE:I2T:S
TATE ON
:ANALYSIS:PANALYZE1:I2T:MEASURE:I2T:S
TATE?
-> :ANALYSIS:PANALYZE1:I2T:MEASURE:I2
T:STATE 1
:ANALysis:PANalyze<x>:I2T:MEASure:I2
T:VALue?
Function Queries an automatically measured value of a
Joule integral.
Syntax
:ANALysis:PANalyze<x>:I2T:MEASure:I2T
:VALue? [{<NRf>}]
<x> = 1 or 2
<NRf> = See the DLM2000 Features Guide for
this information.
Example
:ANALYSIS:PANALYZE1:I2T:MEASURE:I2T:V
ALUE?
-> :ANALYSIS:PANALYZE1:I2T:MEASURE:I2
T:VALUE 10.0000E+03
Description • If the value is immeasurable, the DLM2000
returns “NAN” (not a number).
• The <NRf> is used to specify which iteration of
automated measurement to query the measured
value from.
• If <NRf> is set to 1, the oldest measured value
in the automated measurement memory is
queried.
• If a measured value does not exist at the
specified iteration, the DLM2000 returns “NAN”
(not a number).
• If <NRf> is omitted, the most recent measured
value is queried.
• If cyclic statistical processing for automated
measurement is being executed and <NRf> is
specified, the measured values over a cycle in
iteration <NRf> from the left of the display on
the displayed waveform is queried. If <NRf>
is omitted, the measured values over the last
cycle on the displayed waveform is queried.
:ANALysis:PANalyze<x>:I2T:RANGe
Function Sets or queries the measurement source window.
Syntax
:ANALysis:PANalyze<x>:I2T:RANGe {MAIN
|Z1|Z2}
:ANALysis:PANalyze<x>:I2T:RANGe?
<x> = 1 or 2
Example
:ANALYSIS:PANALYZE1:I2T:RANGE MAIN
:ANALYSIS:PANALYZE1:I2T:RANGE?
-> :ANALYSIS:PANALYZE1:I2T:RANGE MAIN
:ANALysis:PANalyze<x>:I2T:SCALe?
Function Queries all scaling settings.
Syntax
:ANALysis:PANalyze<x>:I2T:SCALe?
<x> = 1 or 2
5.3 ANALysis Group