5-58
IM 710105-17E
:ANALysis:AHIStogram<x>:MODE
Function Sets or queries the target axis of a waveform
histogram.
Syntax
:ANALysis:AHIStogram<x>:MODE {HORizon
tal|VERTical}
:ANALysis:AHIStogram<x>:MODE?
<x> = 1 or 2 (1 only for 2-channel models)
Example
:ANALYSIS:AHISTOGRAM1:MODE HORIZONTAL
:ANALYSIS:AHISTOGRAM1:MODE?
-> :ANALYSIS:AHISTOGRAM1:MODE HORIZON
TAL
:ANALysis:AHIStogram<x>:RANGe
Function Sets or queries the measurement source window
of a waveform histogram.
Syntax
:ANALysis:AHIStogram<x>:RANGe {MAIN|Z
1|Z2}
:ANALysis:AHIStogram<x>:RANGe?
<x> = 1 or 2 (1 only for 2-channel models)
Example
:ANALYSIS:AHISTOGRAM1:RANGE MAIN
:ANALYSIS:AHISTOGRAM1:RANGE?
-> :ANALYSIS:AHISTOGRAM1:RANGE MAIN
:ANALysis:AHIStogram<x>:TRACe
Function Sets or queries the source waveform of a
waveform histogram.
Syntax
:ANALysis:AHIStogram<x>:TRACe {<NRf>|
MATH<x>}
:ANALysis:AHIStogram<x>:TRACe?
<x> = 1 or 2 (1 only for 2-channel models)
<NRf> = 1 to 4 (1 or 2 for 2-channel models)
<x> of MATH<x> = 1 or 2 (1 only for 2-channel
models)
Example
:ANALYSIS:AHISTOGRAM1:TRACE 1
:ANALYSIS:AHISTOGRAM1:TRACE?
-> :ANALYSIS:AHISTOGRAM1:TRACE 1
:ANALysis:AHIStogram<x>:VERTical
Function Sets or queries the vertical range of a waveform
histogram.
Syntax
:ANALysis:AHIStogram<x>:VERTic
al {<NRf>,<NRf>}
:ANALysis:AHIStogram<x>:VERTical?
<x> = 1 or 2 (1 only for 2-channel models)
<NRf>=−4to4(divisions)
Example
:ANALYSIS:AHISTOGRAM1:VERTICAL 0,1
:ANALYSIS:AHISTOGRAM1:VERTICAL?
-> :ANALYSIS:AHISTOGRAM1:VERTICAL 1.0
00E+00,0.000E+00
:ANALysis:PANalyze<x>?
Function Queries all power supply analysis settings.
Syntax
:ANALysis:PANalyze<x>?
<x> = 1 or 2
:ANALysis:PANalyze<x>:I2T?
Function Queries all Joule integral settings.
Syntax
:ANALysis:PANalyze<x>:I2T?
<x> = 1 or 2
:ANALysis:PANalyze<x>:I2T:MATH
Function Sets or queries the on/off status of the Joule
integral waveform display.
Syntax
:ANALysis:PANalyze<x>:I2T:MATH {I2T|O
FF}
:ANALysis:PANalyze<x>:I2T:MATH?
<x> = 1 or 2
Example
:ANALYSIS:PANALYZE1:I2T:MATH I2T
:ANALYSIS:PANALYZE1:I2T:MATH?
-> :ANALYSIS:PANALYZE1:I2T:MATH I2T
:ANALysis:PANalyze<x>:I2T:MEASure?
Function Queries all Joule integral automatic measurement
settings.
Syntax
:ANALysis:PANalyze<x>:I2T:MEASure?
<x> = 1 or 2
:ANALysis:PANalyze<x>:I2T:MEASure:I
2T?
Function Queries Joule integral settings.
Syntax
:ANALysis:PANalyze<x>:I2T:MEASure:I
2T?
<x> = 1 or 2
:ANALysis:PANalyze<x>:I2T:MEASure:I2
T:COUNt?
Function Queries the continuous statistical processing
count of a Joule integral.
Syntax
:ANALysis:PANalyze<x>:I2T:MEASure:I2T
:COUNt?
<x> = 1 or 2
Example
:ANALYSIS:PANALYZE1:I2T:MEASURE:I2T:C
OUNT?
-> :ANALYSIS:PANALYZE1:I2T:MEASURE:I2
T:COUNT 100
5.3 ANALysis Group