EasyManua.ls Logo

YOKOGAWA WT310EH - Page 128

YOKOGAWA WT310EH
143 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
App-22
IM WT310E-01EN
Example 2: Measuring 1 Arms Using the 5 A Measurement Range
Inthisexample,wemeasurethesameinputusingthe5Arange.Whenthemeasuredvalueis1.0000
[A],thereadingerrorandrangeerrorareasfollows:
• Readingerror:1.0000[A]×0.1%=0.001[A]
• Rangeerror:5[A]×0.05%=0.0025[A]
Theerrorincludedin1.0000[A]isthesumofthereadingandrangeerrors,whichis±0.0035[A].This
correspondsto0.35%ofthemeasuredvalue.
The error has increased even though the same current signal as example 1 was measured. As this
example illustrates, using a measurement range that is unnecessarily large for an input signal results
in larger measurement errors. It is import to measure using a measurement range that is appropriate
for the input signal.
Note
If the input signal is not a sine wave and includes distortions and spikes, select a somewhat large
measurement range that would not cause peak over-ranges to occur.
Example 3: Measuring 0.5 Arms Using the 1 A Measurement
Range
Next,wemeasure0.5Ausingthe1Ameasurementrange(thesameasexample1).Whenthe
measuredvalueis0.5000[A],thereadingerrorandrangeerrorareasfollows:
• Readingerror:0.5000[A]×0.1%=0.0005[A]
• Rangeerror:1[A]×0.05%=0.0005[A]
Theerrorincludedin0.5000[A]isthesumofthereadingandrangeerrors,whichis±0.001[A].This
correspondsto0.2%ofthemeasuredvalue.
When we compare this result with that of example 1, we notice the following:
• Thereadingerrorhasbeenreducedinaccordancewiththeinputamplitude.
• Therangeerrorhasnotchanged.
Asaresult,theerroris0.2%,whichisslightlylargerthan0.15%ofexample1.Thisisalsobecause
themeasurementrangeislargerelativetotheinputsignal.Inthiscase,weshouldusethe0.5A
measurement range.
Measurement Error of Active Power
Onthisinstrument,thepoweraccuracyintherangeof45Hzto66Hzis±(0.1%ofreading+0.05%
of range).
Let us calculate the error for the following example.
• Voltagemeasurementrange:150V,measuredvoltage:100.00V
• Currentmeasurementrange:1A,measuredcurrent:0.800A
• Measuredpower:80.00W
• 60Hzsinewaveforbothvoltageandcurrent
• Phasedifferencebetweenthevoltageandcurrentsignals=0°
Power Range
The power measurement range is defined as voltage measurement range × current measurement
range.Inthisexample,thepowermeasurementrangeis150V×1A=150W.Weusethispower
measurement range to calculate the range error.
Thereadingerrorandrangeerrorincludedinthemeasuredpower(80.00W)areasfollows:
• Readingerror:80.00[W]×0.1%=0.08[W]
• Rangeerror:150[W]×0.05%=0.075[W]
Theerrorincludedin80.00[W]isthesumofthereadingandrangeerrors,whichis±0.155[W].This
correspondsto0.19375%ofthedisplayedvalue.
Appendix 5 Measurement Accuracy and Measurement Error

Table of Contents

Related product manuals