1.3 Holding Measured Values and Performing
Single Measurements
Holding Measured Values (HOLD; for procedures, see section 3.1)
You can use the hold operation to stop measuring and displaying the measured data at each data
update interval and hold the display of all measurement function data. The D/A output, communication
output, and other values reflect the held values.
Single Measurement (SINGLE; for procedures, see section 3.2)
While the display is held, the signal is measured once at the data update interval, and then the display
is re-held. If you press SHIFT+HOLD (SINGLE) when the display is not held, measurement restarts
from that point.