IM WT310E-01EN
Standard Integration Mode
Youcansetarelativeintegrationtime(setatimer).IntegrationstartswhenyoupressSTART.When
one of the conditions below is met, integration is stopped, and the integration time and integrated value
are held.
• Thespecifiedtimertimeelapses.
• YoupressSTOP.
• Theintegratedvaluereachesitsmaximumorminimumdisplayablevalue.
Start
Hold
Timer value
Integration
Integrated
Continuous Integration Mode
Youcansetarelativeintegrationtime.IntegrationstartswhenyoupressSTART.Whenthespecified
integration timer time elapses, integration is automatically reset and restarted. Integration repeats until
you press Stop. If either of the conditions below is met, integration is stopped, and the integration time
and integrated value are held.
• YoupressSTOP.
• Theintegratedvaluereachesitsmaximumorminimumdisplayablevalue.
Start
Hold
Hold
Timer
value
Timer
value
Timer
value
Stop
Integrated
Integration
Integration Methods (Equations)
For information on the equations used for integration, see appendix 1.