2-25
IM WT310E-01EN
Measurement Conditions
1
2
3
4
5
6
7
8
App
Index
Averaging Coefficient (Attenuation Constant or Average Count)
You can select the attenuation constant (for exponential averages) or average count (for moving
averages)fromthefollowingoptions.Thedefaultsettingforbothvaluesis8.
8,16,32,and64
Note
You can set separate averaging coefficients for exponential averages and moving averages.
Turning Averaging On and Off
After you set the averaging coefficient (the attenuation constant or average count), select whether to
set averaging to on or oFF. The default setting is oFF.
• on:Averagingstarts,andtheAVGindicatorilluminates.
• oFF:Averagingisstopped.TheAVGindicatorturnsoff.
Note
• Whenyoustartintegration,averagingisdisabled.Evenafterintegrationstopsandisreset,averagingis
not enabled again.
• Inharmonicmeasurement,onlyexponentialaveragingisavailable.Fordetails,seesection6.1.
2.11 Setting Averaging