5-2
IM WT310E-01EN
Standard Integration Mode
Youcansetarelativeintegrationtime(setatimer).IntegrationstartswhenyoupressSTART.When
one of the conditions below is met, integration is stopped, and the integration time and integrated value
are held.
• Thespecifiedtimertimeelapses.
• YoupressSTOP.
• Theintegratedvaluereachesitsmaximumorminimumdisplayablevalue.
Hold
Timer value
value
time
Continuous Integration Mode
Youcansetarelativeintegrationtime.IntegrationstartswhenyoupressSTART.Whenthespecified
integration timer time elapses, integration is automatically reset and restarted. Integration repeats until
you press Stop. If either of the conditions below is met, integration is stopped, and the integration time
and integrated value are held.
• YoupressSTOP.
• Theintegratedvaluereachesitsmaximumorminimumdisplayablevalue.
Hold
Hold
Timer
value
Timer
value
Timer
value
value
time
5.1 Integration Feature