8-3
IM WT310E-01EN
Other Features
8
Item Default setting
GP-IB Address: 1
RS-232 Handshaking method: 0, data format: 0, baud rate: 9600, terminator:
Cr+Lf
Ethernet (option) DHCP: ON
When DHCP is set to OFF
IPaddress:192.168.0.100
Subnetmask:255.255.255.0
Defaultgateway:192.168.0.1
Command mode WT300E
Settings That Cannot Be Reset to Their Factory Default Values
• Communication(GP-IB,RS-232,Ethernet,andcommandmode)settings
Data
• Storedmeasureddata
When you initialize the settings, stored measured data is lost.
• Setupparameters
The setup parameters that are stored in internal memory are retained even if you initialize the
settings.
Note
Only initialize this instrument if you are sure that it is okay for all of the settings to be returned to their default
values. You cannot undo an initialization. We recommend that you save the setup parameters before you
initialize this instrument (see section 7.2).
Initializing Using Communication Commands
Initializationofsetupparametersexecutedthroughcommunicationcommands(*RSTcommand)works
in the same way as the initialization executed through the UTILITY key, as describe above.
Initializing at Power-On
If you turn on the power while holding down SET, this instrument will start up with the default settings.
Continue holding SET until after all the LEDs have lit and turned off (part 2 under “Power-on
Messages” on page 2-10 in the Getting Started Guide, IM WT310E-02EN). All of the setup parameters,
including the communication settings, will be initialized.
The message “CodE.3” will appear to indicate that the settings have been initialized. This is not a
malfunction.
8.2 Initializing Settings