OPERATION
Axiovert 200 Axiovert 200 M Carl Zeiss
B 40-080 e 03/01 3-21
LEFT / RIGHT key for the motorized switching of the Sideport turret (3-13/13)
− Key to switch the beam splitting ratio for Sideport (doc) right, Sideport (doc) left and VIS (visual
observation through the binocular).
− Five motorized configurations are possible:
LSM connector SP L and Sideport 60 R
LSM connector BP and Sideport 60 L
LSM connector BP and FCS connector SPL
Safety device LSM / PASCAL
Change to beam path mot. LSM
− Key pressed briefly: current position is shown in the LCD display for 4 s (or until the next press
of the key);
without LCD display, the position is approached immediately
key pressed again: target position is displayed and approached
− The set beam path is displayed depending on the VIS / Frontport / Baseport setting, i.e. the actual
intensities at the ports are displayed (in %).
Key for the motorized switching of the VIS / Frontport / Baseport slider (3-13/14)
− Key to switch the beam path between Frontport (FP), Baseport (BP) and VIS (visual observation
through the binocular).
− The display in the LCD display depends on the position of the Sideport turret.
− The following is displayed: Binocular, Baseport, Sideport
− Empty positions are not approached or displayed unless a different setting has been made in the
AXIOSET software program.
FCT key (3-13/15)
− The AXIOSET software program permits any required CAN/SUB components to be assigned to this
key.
− Unless the key has been assigned otherwise, the instrument is brought to a basic position (Home or
Default position) after activation.
Basic position:
Optovar 1x (pos 1)
HAL = 3 V
SP 100 % VIS (pos 1)
BP 100 % VIS (pos 2)
condenser aperture fully opened
condenser turret BF (pos 1)
reflector turret on transmission
FL shutter closed
Key "Set LM" (3-13/18)
− briefly pressed: Light Manager values are saved
pressed long (> 2 s): activation of Set mode