Switching on
To switch on the device, press the power switch (6, Fig. 9) followed by
the Standby button (3, Fig. 6). The i.Profiler
® plus
software is
automatically launched and the Patient window will open.
After switching on, a function test of the measurement functions on the
test eye pair must be performed pursuant to the instructions in the
i.Profiler
® plus
Software description.
CAUTION - RISK OF MEASURING ERRORS
The calibration of the test eye pair should be checked each day prior to
measurements on patients. The measured values can be printed out and
filed for documentation purposes. If the values obtained from the test eye
pair are not within the given tolerances, no patient measurements may be
taken! The device must be taken out of service immediately (see section
Shutting down, page 45).