Maintenance and care
000000-1520-928_GA_EN_021013
52
Maintenance
Replacing the print paper roll
WARNING - RISK OF ELECTRIC SHOCK
Disconnect the power cable from the power supply before replacing the
printing paper.
Open the paper compartment at the right hand side of the device
(8, Fig. 9, page 22).
Remove the empty paper roll and insert a new roll according to the
drawing on the inside of the compartment.
Pull out the end of the print paper roll so that it remains visible when the
cover has been closed.
Close the printing paper compartment.
Care and cleaning
WARNING - RISK OF ELECTRIC SHOCK
Prevent moisture from penetrating the instrument or keyboard. Disconnect
the power cable from the power supply before cleaning or disinfecting the
device.
CAUTION - RISK DUE TO CROSS-CONTAMINATION
Parts with which the patient has come into contact during the examination
(chin rest, forehead rest) should be cleaned with a disinfectant approved for
the purpose. These parts are resistant to wiping off with cleaning agents
categorized as "low" (e.g. suds, quaternary ammonium compounds) and
"intermediate" (e.g. alcohol, Javel water, iodine); classification pursuant to:
Disinfectants and activity spectrum according to the Center for Disease
Control and Prevention, Atlanta, USA).