Electromagnetic compatibility
000000-1520-928_GA_EN_021013
Recommended separation distances between portable and mobile RF communications equipment and the
i.Profiler
® plus
The i.Profiler
® plus
is intended for use in an electromagnetic environments in which related RF disturbances are
controlled. The customer or the user of the i.Profiler
® plus
can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the i.Profiler
® plus
as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
output power of
transmitter
P [W]
Separation distance, according to frequency of transmitter
d [m]
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
d =
d =
d =
0.01 0.117 0.117 0.233
0.1 0.370 0.370 0.737
1 1.170 1.170 2.330
10 3.700 3.700 7.368
100 11.700 11.700 23.300
For transmitters rated at a maximum output power not listed above, the recommended separation distance d
in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, in which P is
the maximum output power of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 and 800 MHz the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply to all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.