Section 10.0
MAN 06–0017H
ZIEHM QUANTUM User’s Manual Page 164 of 220
10.13.2.2 DELETE ALL PERMISSIONS:
Administration Permission can also delete all but the Administrative person that is logged into the
screen.
• Once the Log In screen is visible on the touch panel press the DELETE ALL button.
• If you wish to continue Press yes in the message box
The system will remove all but the present logged in administrative person.
10.13.2.3 SAVE AND CLEAR EXPORT LOG FILES:
The System creates Log In files for each export of images
from the archive and USB, DICOM functions.
The Administrator has permission to Copy the Log file to a
USB memory device and to delete the log file.
CAUTION
The log file should be copied once a week to keep it from getting too large. The system
will stop acquiring the log files when the file reaches it maximum capacity. There is no
alert for this so the Administrator is encouraged to copy and then reset the Log file once
a week.
The Log files can be found at the bottom of the Configure screen center and will only be visible if
the Administrator has logged into the system. See example of the configuration screen above in
this section of the manual.
User configurations
The Administrator and the User may select the start up default settings for several of the image
operations, such as, set the NR to start at low, med, or high. Select invert, image mirror
operations, Last image Noise reduction “LIH”, etc.
See Configuration screen below for selections and location of the buttons.