Section 5.0
MAN 06–0017H
ZIEHM QUANTUM
User’s Manual Page 53 of 220
Fig 5.3: C-Arm Stand Steering (schematic representation, view from above)
5.3 MOVEMENT OF THE C-ARM, MECHANICAL BRAKES
The C-Arm can be moved in virtually any plane. For each movement, a separate brake is
available, allowing you to secure the C-Arm in any position.
To release the corresponding brake, turn it counter-clockwise.
The limit stop of the brake handles can be adjusted individually. To do so, lift the spring-locking
brake handle, turn it until reaching the desired limit stop position and release it to re-engaged it at
the new position.
CAUTION
Before moving the C-Arm, make sure that there is nobody within its range of
movement.