Section 4.0
MAN 06–0017H ZIEHM QUANTUM User’s Manual Page 38 of 220
1. To avoid damage to the equipment by condensation of moisture resulting from temperature
differences, make sure that all parts of the equipment have reached room temperature
before turning the equipment on.
2. Remove the outside of the shipping container and inspect for damage.
3. Lower the unloading ramp into position.
4. Remove the metal brackets that hold the C-Arm mobile stand to the pallet. Take care not to
scratch the mobile stand.
5. Remove the two straps that hold the monitors and take the monitors off the pallet.
6. Remove all of the brackets from the pallet.
7. Loosen the Wig-Wag knob and straighten the C-Arm. This will keep the C-Arm from
tipping over as it is removed from the pallet.
8. Lift the lever located in the center of the control panel (the hand brake) and carefully move
the mobile stand off the pallet and onto the floor.

Table of Contents