FEM630 | ELECTROMAGNETIC FLOWMETER | OI/FEM630-EN REV. A
8 ...Operation
...Parameter descriptions
Menu: Sensor Setup
Note
Menu only available if function ‘Backwards Comp.’ has been ordered and activated (see menu ’Device Setup/...Transmitter/...
Feature Settings‘).
If the function Backwards Comp. is activated, the transmitter can also be used with older transducers.
Setting the parameters in accordance with the specification on the name plate of the older sensor.
Software history
In accordance with NAMUR recommendation NE53, ABB offers a transparent and traceable software history.
Damage to components!
Damage to the flowmeter sensor due to incorrect setting of the excitation current possible.
NOTICE
Menu / parameter Description
Sensor Setup
Sensor Type
Select the sensor type:
• Standard QmaxDN: For ProcessMaster, HygienicMaster.
Meter Size
Setting the nominal diameter in accordance with the value provided on the name plate of the flowmeter
sensor.
Se Span Setting the span in accordance with the value provided on the name plate of the flowmeter sensor.
Setting the zero point in accordance with the value provided on the name plate of the flowmeter sensor.
Line Freq
Excitation Freq.
Setting the excitation frequency in accordance with the value provided on the name plate of the flowmeter
sensor.
Set Point Curr.
Adjustment of the excitation current of the sensor coils.
For all other sensors contact ABB Service.
Pre. Amp
Selection whether a preamplifier exists in the flowmeter sensor or not
preamplifier.
Calibration flag
Sensor TFE Function This activates or deactivates full pipe detection (TFE = complete fill electrode).
Design Issue date Type of change Description Ordering number
First publication –
Bug fixing Piston pumps filter
Bug fixing Integrated Polish language
Bug fixing PROFIBUS DP® and Modbus® integrated. New bootloader