38
FEM630 | ELECTROMAGNETIC FLOWMETER | OI/FEM630-EN REV. A
Digital output 41 / 42, 51 / 52
Can be configured as pulse, frequency or binary output via
on-site software.
Current output V1 / V2, V3 / V4 (plug-in module)
Up to two additional plug-in modules can be implemented via
the ‘Passive current output (red)’ option module.
The plug-in module can be used in slot OC1 and OC2.
Note
Terminals 42 / 52 have the same potential. Digital
outputs DO 41 / 42 and DO 51 / 52 are not electrically
isolated from each other. If an additional electrically
isolated digital output is required, a corresponding plug-
in module must be used.
If you are using a mechanical counter, we recommend

frequency of f
max

6 ...Electrical connections
...Pin assignment
Digital output 41 / 42, 51 / 52 passive as a pulse or frequency
output
Passive digital output 51 / 52 as binary output
Current output V1 / V2, passive
Current output V3 / V4, passive
Permissible source voltage U
q
for passive outputs in relation to load
resistance R
B
where I
max
= 22 mA. = Permissible range
A
B
A
B
Figure 41 
B

Figure 42 
B

Figure 43 Source voltage for passive outputs
G11597-02
A
16...30 VDC
B
R
B
U
CE
I
CE
R
B
U
CE
I
CE
41+
51+
IE
42-/
52-
41+
51+
IE
42-/
52-
R
B
R
B
012345
16...30 V DC
012345
R
B
Pulse / frequency output (passive)
Terminals

Output ‘closed’


CEL


CEL

Output ‘open

CEH


CEH

f
max

Pulse width

Passive current output
Terminals

Output signal

Load R
B

Source voltage U*

Measuring error

Resolution

Binary output (passive)
Terminals

Output ‘closed’

CEL


CEL

Output ‘open

CEH


CEH

Switching function
Parameterization possible.
See Menu: Input/Output 
G11897-02
A
B
V1+
IE
V2-
R
B
OC1
V3+
IE
V4-
R
B
OC2
G10323-02
0
0
100
200
300
400
500
600
5101520253035
RB [Ω]
Uq [V]
*The source voltage U
q
is dependent of the load R
B
and must be placed in
an additional area.