100  FEP630, FEH630 ELECTROMAGNETIC FLOWMETER  |  OI/FEP630/FEH630-EN REV. D   
 
… 8  Operation 
… Parameter descriptions 
 
Diagnostics / ...Diagnosis Control / ...Diagnosis Conductiv* 
Conductivity On/Off  Activate the conductivity diagnostic function. 
Default setting: Off 
Note 
Gas bubble detection can be used in the nominal diameter range of DN 10 to 300. 
For additional information, see Extended diagnostic functions on page 118. 
Conductivity[µS/cm]  Indicator of the measured conductivity in μS/cm. 
Adj. Cond. Value  Conductivity must be set in accordance with the conditions on-site. 
Measure the conductivity using a conductivity meter on-site and enter the measured value here. 
Limits: 5 to 20000 μS/cm 
Cond. Iout Min Value  The conductivity value is available as a 4 to 20 mA-output (option card). 
Set the 4 mA and 20 mA value which correspond to the upper and lower range of the conductivity value. 
Cond. Iout Max Value 
Cond.Min Alarm Value  Set the alarm for minimum and maximum conductivity. In the case of down-scale, an alarm is triggered. 
Limits: 5 to 20000 μS/cm 
Cond.Max Alarm Value 
Elec. Imp. E1-GND  Electrical impedance E1-GND. 
Current impedance between electrode E1 and GND (ground potential). 
Elec. Imp. E2-GND  Electrical impedance E2-GND. 
Current impedance between electrode E2 and GND (ground potential). 
 
Diagnostics / ...Diagnosis Control / ...Diagnosis SIL** 
SIL On/Off  For information purposes only. 
SIL devices are delivered ex works as SIL devices. There is no special SIL mode to activate 
*  The menu is only available if the Expert Diagnosis function is activated. See also the ‘Device Setup\...Transmitter\...Feature Settings’ menu. 
**  Menu only available if SIL diagnostic function is activated. See also the ‘Device Setup\...Transmitter\...Feature Settings’ menu.