EasyManua.ls Logo

ABB ProcessMaster FEP630 series - Page 117

ABB ProcessMaster FEP630 series
152 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
FEP630, FEH630 ELECTROMAGNETIC FLOWMETER | OI/FEP630/FEH630-EN REV. D 117
Error no. / Range Error text Current output Digital output Pulse output LCD display Error maskable?
S047.015 /
operation
Pulse output is cut off.
Wrong config. Check pulse out
configuration.
Current value - no
change.
Current value - no
change.
Largest possible
pulse rate
Current value - no
change.
Menu ‘...Group
Masking’.
S046.000 /
operation
Mass flowrate exceeds limits.
Check flowrate and alarm limits.
Current value - no
change.
Alarm, if DO as ‘Qm
Massflow Max’ Or
‘Qm Massflow Min’
is configured.
Current value - no
change.
Current value - no
change.
Menu ‘...Group
Masking’.
S044.001 /
operation
Volume flowrate exceeds limits.
Check flowrate and alarm limits.
Current value - no
change.
Alarm, if DO as ‘Qv
Volumeflow Max’ Or
‘Qv Volumeflow Min’
is configured.
Current value - no
change.
Current value - no
change.
Menu ‘...Group
Masking’.
S041.034 /
electronics
DC feedback regulation.
Check conditions of application.
Call service.
4 mA (0 % flow) No Answer. 0 Hz 0 % Menu ‘...Group
Masking’.
S040.031 /
electronics
Coil Inductance alarm.
Call service.
Current value - no
change.
No Answer. Current value - no
change.
Current value - no
change.
Menu ‘...Group
Masking’.
M038.009 /
electronics
Sensor memory defective.
Mem. or connect. defective.
Replace memory.
Current value - no
change.
No Answer. Current value - no
change.
Current value - no
change.
Menu ‘...Group
Masking’.
M037.014 /
Electronics
NV chips defect on
Motherboard.
Defective MB. Replace MB. Call
Service.
M032.022 / Config.
Curr.Out 31 / 32 not calibrated.
Call Service.
M031.023 / Config.
Curr.Out V1 / V2 not calibrated.
Replace Current Option Card.
Call Service.
Current value - no
change.
Current value - no
change.
Current value - no
change.
Current value - no
change.
Menu ‘...Group
Masking’.
M030.024 /
Config.
Curr.Out 31 / 32 not calibrated.
Call Service.
M028.007 /
Config.
Display value is < 1600 h at
Qmax.
Change mass Unit or vol. Unit
for Totalizer.
Current value - no
change.
No Answer. Current value - no
change.
Current value - no
change.
Menu ‘...Group
Masking’.
M026.004 /
Config.
Maintenance interval is reached.
Perform maintenance.
M024.008 /
Config.
Device not calibrated.
Call Service.
High Alarm or Low
Alarm, depending on
parameter ‘Curr.Out
at Alarm’.
Current value - no
change.
Current value - no
change.
Current value - no
change.
Menu '...Group
Masking’.
M020.027 /
Electronics
Communicat. Card not
responding
Current value - no
change.
Current value - no
change.
Current value - no
change.
Current value - no
change.
Menu ‘...Group
Masking’.
Change from one to tw o columns

Table of Contents

Related product manuals