116 FEP630, FEH630 ELECTROMAGNETIC FLOWMETER | OI/FEP630/FEH630-EN REV. D
… 9 Diagnosis / error messages
… Overview
Error no. / Range Error text Current output Digital output Pulse output LCD display Error maskable?
S065.044 /
operation
Inhouse temp. alarm.
Reduce ambient temperature.
Current value - no
change.
No Answer No Answer Current value - no
change.
S064.041 /
operation
EPD alarm.
Secure pipe is completely filled.
Alarm - configured
as in menu ‘Curr. at
EPD Alarm’.
Alarm, if DO as
‘Logic / Alarm Signal
/ EPD Alarm’ is
configured
0 Hz 0 % Menu ‘...Group
Masking’.
S063.040 /
operation
TFE alarm.
Secure pipe is completely filled.
Alarm - configured
as in menu ‘Curr. at
TFE Alarm’.
Alarm, if DO as
‘Logic / Alarm Signal
/ TFE Alarm’ is
configured
Current value - no
change.
Current value - no
change.
Menu ‘...Group
Masking’.
S062.039 /
operation
Sensor temp. limits alarm.
Change limits or change fluid
temperature.
Current value - no
change.
No Answer Current value - no
change.
Current value - no
change.
Menu ‘...Group
Masking’.
S061.038 /
operation
Conductivity limits alarm.
Change limits or Check
application.
Current value - no
change.
Alarm, if DO as
‘Logic / Alarm Signal
/ Conductivity’ is
configured
Current value - no
change.
Current value - no
change.
Menu ‘...Group
Masking’.
S060.037 /
operation
Gas bubble alarm.
Check conditions of application.
Current value - no
change.
Alarm, if DO as
‘Logic / Alarm Signal
/ Gas bubble Alarm’
is configured
Current value - no
change.
Current value - no
change.
Menu ‘...Group
Masking’.
S052.016 /
operation
Curr.Out 31 / 32 is saturated.
CO process value out of range.
Adapt Qmax.
Alarm - configured
as in menu
‘„Curr.Out > 20.5mA’.
Current value - no
change.
Current value - no
change.
Current value - no
change.
Menu ‘...Group
Masking’.
S051.017 /
operation
Curr.Out V1 / V2, V3 / V4
saturated.
CO process value out of range.
Adapt Qmax.
S049.019 /
electronics
Option Card 1 com error.
Defective Card. Check Card 1.
Call Service.
Does not react
anymore
Current value - no
change.
Current value - no
change.
Current value - no
change.
Menu ‘...Group
Masking’.
S048.020 /
electronics
Option Card 2 com error.
Defective Card. Check Card 2.
Call Service.
Does not react
anymore
Current value - no
change.
Current value - no
change.
Current value - no
change.
Menu ‘...Group
Masking’.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ABB ProcessMaster FEP630 series and is the answer not in the manual?

ABB ProcessMaster FEP630 series Specifications

General IconGeneral
BrandABB
ModelProcessMaster FEP630 series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals