58  FEP630, FEH630 ELECTROMAGNETIC FLOWMETER  |  OI/FEP630/FEH630-EN REV. D   
 
… 7  Commissioning 
… Parameterization via the menu function Easy Setup 
 
Easy Setup   
Pulse Width 
 
30.00 ms
Next  Edit
 
22. Use   to call up the edit mode. 
23. Use   /   to select the desired pulse width for the pulse 
output.. 
24.Confirm the selection with  . 
 
Easy Setup   
Damping 
 
30.00 ms
Next  Edit
 
25.Use   to call up the edit mode. 
26.Use   /   to set the desired damping. 
27. Confirm the selection with  . 
 
Easy Setup   
Dig.Out 41/42 Mode 
 
Pulse
Next  Edit
 
28.Use   to call up the edit mode. 
29.Use   /   to select the desired operating modeOff, 
Logic, Pulse, Frequencyfor the digital output. 
30.Confirm the selection with  . 
 
Easy Setup   
Curr.Out at Alarm 
 
High Alarm
Next  Edit
 
31. Use   to call up the edit mode. 
32. Use   /   to select the desired alarm mode. 
33.Confirm the selection with  . 
 
 
Easy Setup   
Low Alarm 
 
3.500 m
Next  Edit
 
34.Use   to call up the edit mode. 
35.Use   /   to set the desired current for Low Alarm. 
36.Confirm the selection with  . 
 
Easy Setup   
High Alarm 
 
21.800 mA
Next  Edit
 
37. Use   to call up the edit mode. 
38.Use   /   to set the desired current for High Alarm. 
39.Confirm the selection with  .