FEP630, FEH630 ELECTROMAGNETIC FLOWMETER  |  OI/FEP630/FEH630-EN REV. D  59 
 
Zero point adjustment of the flowmeter 
 
Note 
Prior to starting the zero point adjustment, make sure that: 
•  There is no flow through the sensor (close all valves, shut-
off devices etc.) 
•  The sensor is completely filled with the medium to be 
measured 
 
Easy Setup   
System Zero 
 
       
Next  O
 
•  Use   to start automatic adjustment of the zero point for 
the system. 
 
Menu 
 
 
Easy Setup 
 
Exit  Select
 
Once all parameter have been set, the main menu appears again. 
The most important parameters are now set. 
 
40.Use   to switch to the process display. 
 
 