19
Control Solution Test
A control solution test should be performed when you are not sure of your results and want to
conrm that your meter and test strips are working properly.
IMPORTANT:
• Control solution results should fall within the control solution range printed on the test
strip vial.
• Do not use control solution past its expiration date. Discard control solution 3 months
after opening or on the expiration date printed on the bottle, whichever comes rst.
(Example: open April 15, discard July 15; write the discard date on the side of the bottle.)
• The control solution range is a target range for control solution only, not for your blood
glucose level.
• Use only FreeStyle control solution with this meter.
• Replace the cap securely on the bottle immediately after use.
• Do not add water or other liquid to control solution.
• Contact Customer Service for information on how to obtain control solution.