21
3
Compare the control solution result to the range printed on the test strip vial. The result
should fall within the range.
Notes:
If you see an error message on the screen, your meter will explain what it means,
its possible causes, and what to do next.
The result will be marked as a control solution test in your logbook. It will not be
included in your charts and reports.
You may also mark a control solution test after it is performed by following these
steps:
a. At the Home screen, tap Logbook .
b. Tap your most recent test. You can only modify this result.
c. Tap the down arrow. Then tap add or edit notes.
d. Tap the option for control solution test. Then tap OK.
e. You can now see your most recent result marked as a control solution test.
Out of Range Control Solution Results
Repeat the test if control solution results are outside the range printed on the test
strip vial.
Stop using the meter if control solution results are consistently outside the range
printed on the test strip vial. Contact Customer Service.