m-PIMA™ ANALYSER USER GUIDE
63
9
Error Code Acon
21111-21115, 21120
Make sure that the sample capillary is
completely lled. Keep analyser door closed
during test run.
Repeat test with new cartridge.
Contact your local distributor or your regional
Technical Support if the error persists.
21501
Repeat test with new cartridge.
Make sure that the sample capillary is
completely lled.
21502
Too low atmospheric pressure. Do not connue
with tesng unl you have reached altudes that
are within specicaons.
21503
Too high atmospheric pressure. Do not connue
with tesng unl you have reached altudes that
are within specicaons.
23020
Cartridge or sample related errors.
Please refer to the test cartridge guides PI-
m-PIMA-04 or PI-m-PIMA-05 for further
informaon about sample storage and
handling.
23028
Unexpected amplicaon curve, e.g. due to
individual sample properes.
Repeat test with new cartridge.
Contact your local distributor or your regional
Technical Support if the error persists.