EasyManuals Logo

Abbott m-PIMA User Manual

Abbott m-PIMA
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #8 background imageLoading...
Page #8 background image
1
1 INTRODUCTION
8
Declaraon of Conformity
The m-PIMA™ Analyser complies with Direcves 98/79/EC and 2011/65/EU
and was tested according to IEC 61010-1, IEC 61010-2-101, IEC 61326-1 and
IEC 61326-2-6.
Unpacking, Installaon and Transport
Unpack the m-PIMA™ Analyser and place it safely on a clean, at, level and
stable surface. Inspect for any obvious sign of damage. Report any damage
immediately to your local distributor.
Allow 5 10 cm space on the back of the instrument to ensure free venlaon.
Please use the device handle to li the instrument, considering the instrument
weight. If you transported or moved the m-PIMA™ Analyser to areas with
dierent climac condions, allow temperature equilibraon for at least
30 minutes before turning on the instrument.
Note: Keep the original m-PIMA™ Analyser box in case you need to transport
the instrument. Throw the main power switch on the back of the
instrument to posion ( O ) to prevent accidental start up during
transport.
Calibraon
The m-PIMA™ Analyser is factory calibrated and does not require any further
calibraon. However, upon switching on the m-PIMA™ Analyser performs an
internal roune to verify calibraon and pre-set instrument parameters.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Abbott m-PIMA and is the answer not in the manual?

Abbott m-PIMA Specifications

General IconGeneral
BrandAbbott
Modelm-PIMA
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals