Installation
Installation Warnings
3. Installation
3.1 Installation Warnings
Always make sure that the power switch on the RTCA Analyzer is in the OFF
position, and remove the power cord from the back of the instrument before installing
or relocating the instrument.
Do not operate the instrument in an environment where potentially damaging liquids
or gases are present.
Please do NOT put the RTCA Control Unit and RTCA Analyzer in the incubator!
Connect the RTCA Analyzer to a grounded AC supply line using the supplied cord. is
is not only for safety purposes, but also for better noise performance of the RTCA SP
Instrument. While a stand-alone RTCA SP Instrument can work with a two-pronged AC
power source, but within a complicated electrical environment, it is highly advisable to
use a well-grounded power socket, otherwise there may be a risk of electrical shock, and/
or the performance of the RTCA SP Instrument may be a ected.
A er installing an RTCA SP Station in an incubator, allow at least TWO hours before
to operate or switch on the system. is allows the Station to reach equilibrium with
the incubator environment and reduces the risk of condensation-related problems.
Do not touch or loosen any screws or parts other than those speci cally designat-
ed in the instructions. Doing so might a ect instrument performance and void the
instrument warranty.
3.2 Unpacking
e RTCA Analyzer and RTCA SP Station are packed in a specially designed box.
Retain the box and packing materials. If the system needs to be relocated or returned
for repair, it should be always transported in the packaging provided. It is essential
that you report any suspected damage immediately to your local ACEA representative
and to the shipping agent before accepting the product.
e RTCA Analyzer weighs approximately 7.4 kg and should be li ed with care,
taking proper precautions to avoid injury.
A er examining the box, place it on a at surface in the upright position. Open the top
of the box and remove the top packing material covering the RTCA Analyzer, and li the
RTCA Analyzer, along with the packing materials around the edges, up and out of the
shipping box. Remove the packing materials from the ends of the instrument and set the
instrument down carefully.
RTCA SP Instrument Operator’s Manual
3.1 Installation Warnings
Always make sure that the power switch on the RTCA Analyzer is in the OFF
position, and remove the power cord from the back of the instrument before installing
or relocating the instrument.
Do not operate the instrument in an environment where potentially damaging liquids
Please do NOT put the RTCA Control Unit and RTCA Analyzer in the incubator!
Connect the RTCA Analyzer to a grounded AC supply line using the supplied cord. is
is not only for safety purposes, but also for better noise performance of the RTCA SP
Instrument. While a stand-alone RTCA SP Instrument can work with a two-pronged AC
power source, but within a complicated electrical environment, it is highly advisable to
use a well-grounded power socket, otherwise there may be a risk of electrical shock, and/
or the performance of the RTCA SP Instrument may be a ected.
A er installing an RTCA SP Station in an incubator, allow at least
to operate or switch on the system. is allows the Station to reach equilibrium with
the incubator environment and reduces the risk of condensation-related problems.
Do not touch or loosen any screws or parts other than those speci cally designat-
ed in the instructions. Doing so might a ect instrument performance and void the
e RTCA Analyzer and RTCA SP Station are packed in a specially designed box.
e RTCA Analyzer and RTCA SP Station are packed in a specially designed box.
e RTCA Analyzer and RTCA SP Station are packed in a specially designed box.
e RTCA Analyzer and RTCA SP Station are packed in a specially designed box.
Retain the box and packing materials. If the system needs to be relocated or returned
Retain the box and packing materials. If the system needs to be relocated or returned
Retain the box and packing materials. If the system needs to be relocated or returned
Retain the box and packing materials. If the system needs to be relocated or returned
for repair, it should be always transported in the packaging provided. It is essential
for repair, it should be always transported in the packaging provided. It is essential
for repair, it should be always transported in the packaging provided. It is essential
for repair, it should be always transported in the packaging provided. It is essential
that you report any suspected damage immediately to your local ACEA representative
that you report any suspected damage immediately to your local ACEA representative
that you report any suspected damage immediately to your local ACEA representative
that you report any suspected damage immediately to your local ACEA representative
and to the shipping agent before accepting the product.
e RTCA Analyzer weighs approximately 7.4 kg and should be li ed with care,
e RTCA Analyzer weighs approximately 7.4 kg and should be li ed with care,
e RTCA Analyzer weighs approximately 7.4 kg and should be li ed with care,
taking proper precautions to avoid injury.
A er examining the box, place it on a at surface in the upright position. Open the top
of the box and remove the top packing material covering the RTCA Analyzer, and li the
RTCA Analyzer, along with the packing materials around the edges, up and out of the
shipping box. Remove the packing materials from the ends of the instrument and set the
instrument down carefully.