Technical Specications
ACIST RXi System User’s Guide | 97901700-001,01 2019-09 English
Other Specications
Transportation and Storage Requirements
Requirement Range - System Range - Catheter
Ambient Temperature -20° to 140°F
-29° to 60°C
-20° to 140°F
-29° to 60°C
Relative Humidity 5 to 95%
Non-condensing
5 to 85%
Non-condensing
Atmospheric Pressure 7 to 15 psi
50 to 106 kPa
11 to 15 psi
77 to 106 kPa
Operating Environment Requirements
Requirement Range - System Range - Catheter
Ambient Temperature 64° to 86°F
18° to 30°C
64 to 86°F
18 to 30°C
Relative Humidity 10 to 95%
Non-condensing
10 to 95%
Non-condensing
Atmospheric Pressure 11 to 15 psi
77 to 106 kPa
11 to 15 psi
77 to 106 kPa
Weight
Component Weight
System console (with power supply
and cables)
20 lb (9.1 kg)
Dimensions
Component Dimensions
System console (without power supply
and cables)
Depth: 3.5 in (8.9 cm)
Width: 10.7 in (27.2 cm)
Height: 9.2 in (23.4 cm)
Other
Specications