Basic Operating Procedures
ACIST RXi System User’s Guide | 31901700-001,01 2019-09 English
Verify that the Aortic Pressure is Zero
Zero Aortic Pressure
CAUTION
Ensure aortic pressure is correctly set to zero on the RXi System. Always verify that the Pa
value is zero.
If the system has been set up as a mobile device, you must zero the aortic
pressure signal before each case.
With the hemodynamic monitor still reading zero, re-zero the RXi aortic
pressure if necessary. If using the Mobile setting for the Lab ID (as described
on page 55):
• Press the Zero Aortic button.
• If zeroing was previously performed, the system conrms the action by
displaying the following message:
Do you want to zero the aortic pressure?
• Press OK.
Note
If the Ao Interface Box is connected to the RXi System, refer to Appendix A for additional
information on zeroing.
Verify that
the Aortic
Pressure is
Zero