ACIST RXI - Daily Cleaning

ACIST RXI
124 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance
ACIST RXi System User’s Guide | 77901700-001,01 2019-09 English
Daily Cleaning
Daily cleaning is required for optimal operation.
WARNING
To prevent electric shock, always turn o power to the system and disconnect the power
cord from the main power source before performing any cleaning or service procedure.
Never touch any pins on connectors or cables that have become disconnected from a
system while power is on.
Never immerse any system component in water or in a cleaning solution.
CAUTION
Always wear protective gloves and eye protection when cleaning or decontaminating
the system.
CAUTION
Never allow oil or materials containing oil to come in contact with any part of the
system console.
Never spray or pour any liquid directly onto the power supply, system console, or control
panel.
Do not use any sharp objects or apply excessive pressure to any component.
Discard all cleaning materials in accordance with all local, state, and federal regulations,
codes, and directives.
1. Turn o power to the system.
2. Disconnect the power cord from the main power source.
3. Clean the system console with a so cloth moistened with warm water. Pay
special attention to the system console and power supply where uids may
have dripped.