Installation Precautions
18 | ACIST RXi System User’s Guide 901700-001,01 2019-09 English
Installation
ACIST Medical Systems recommends that you not remove package contents
until instructed to do so. is prevents damaging or misplacing any
equipment or hardware.
CAUTION
To prevent equipment damage, accessory equipment connected to the analog or
digital interfaces on the RXi System must be certied to the IEC standard UL/IEC 60601-
1 for medical equipment. Any person who connects equipment to the input or output
signal port is conguring a medical system. Therefore, this person is responsible for
ensuring that the system complies with the requirements of UL/IEC 60601-1. If in doubt,
consult ACIST Medical Systems Technical Services or your local ACIST representative for
assistance.
CAUTION
Mains power quality should be that of a typical commercial or hospital environment. To
ensure that the RXi System remains operational during power mains interruptions, ACIST
recommends that an uninterruptible power supply or battery backup be used to power
the system.
Before beginning installation, perform the following steps:
• Review possible mounting locations to ensure that the console
touchscreen can be easily reached and viewed by the user.
• Verify that the mounting plate on the mounting post is securely
fastened to the console. Tighten the screws if necessary.
• Ensure that the selected location of the ACIST RXi System is
appropriate for the user’s needs.
• Review the access to the AC main power source. e power supply
must be placed close to an outlet under the bed. Its AC line cord
measures approximately 8 feet (2.5 meters).
• Ensure the host hemodynamic system has an available Pa output of 100
mmHg/V or 20 mmHg/V signal.
Installation
Precautions