EasyManua.ls Logo

ACIST RXI - Perform Pa Scaling

ACIST RXI
124 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installation
ACIST RXi System User’s Guide | 27901700-001,01 2019-09 English
Installation Procedures
Perform Pa Scaling
Select the Waveform Settings option on the ACIST Main Menu, which
provides a submenu of waveform conguration options. From this submenu,
select Pa Scaling.
Aer selecting Pa Scaling, the following screen is displayed.
Aer zeroing the aortic pressure, Pa scaling must be performed if the value
on the hemodynamic system is dierent from the Pa value that appears on
the RXi touchscreen. If the two values are equal or within ±1 mm Hg, it is not
necessary to perform this procedure.
Note
The signal from the hemodynamic system must be something other than zero before
this procedure can be performed.
If the transducer has a backplate that creates a steady test pressure, use this
to articially apply a steady test pressure of 100 mm Hg. is is preferable to
using a live pressure. However, if unable to send a steady pressure, getting the
two values to match can be achieved, but may require more eort.
Use the + and − buttons on the RXi screen to adjust the signal from the
hemodynamic system (1.030 was used in this example) to match the RXi Pa
value displayed on the hemodynamic system screen. When these are equal,
press OK.

Table of Contents

Related product manuals