ACIST RXi System User’s Guide | 75901700-001,01 2019-09 English
Cleaning and Maintenance
7 Maintenance
Routine maintenance is required to keep the ACIST RXi® System in good
working order. is section provides thorough instructions for performing the
following:
daily maintenance items
monthly maintenance and inspection items
Prevent Equipment Damage
To prevent equipment damage, adhere to the following requirements:
Do not clean inside the Navvus MicroCatheter connection port
without the proper cleaning tools. Refer to Optional Fiber Optic
Cleaning Process on page 78.
Use only a so cloth dampened with water to clean the system. Do not
use detergents, cleaning solutions, isopropyl alcohol, or steam to clean
the system.
Never spray or pour any liquid directly onto the power supply or the
system console.
Never allow oil or materials containing oil to come in contact with any
part of the system console.
Do not use any sharp objects.
Discard all cleaning materials in accordance with all local, state, and
federal regulations, codes, and directives.