Warnings
6 | ACIST RXi System User’s Guide 901700-001,01 2019-09 English
Warnings, Cautions, and Symbol Denitions
Improper technique may damage a the Navvus MicroCatheter.
Exercise care in handling the microcatheter during a procedure to
reduce the possibility of accidental breakage or kinking.
Do not disturb pressure sensing system elements during equalization
or FFR recording.
If a broken ber optic is detected within the Navvus MicroCatheter,
a visual and auditory technical alarm condition is triggered. e
defective microcatheter, along with the guide catheter, must be
immediately removed and replaced.
If the system is set to Stationary mode and you move the system from
one lab to another, you must manually zero the RXi aortic pressure
aer you zero the hemodynamic monitoring system or an incorrect
FFR or resting Pd/Pa value may result.
To prevent electric shock, always turn o power to the system and
disconnect the power cord from the main power source before
performing any cleaning or service procedure.
Never touch any pins on connectors or cables that have become
disconnected from a system while power is on.
Never immerse any system component in water or in a cleaning
solution.
is equipment is not suitable for use in the presence of ammable
anesthetic agents.
e RXi system should not be used in rooms containing magnetic
resonance imaging (MRI) equipment.