Chapter 3 Configuring settings
48
OsmoTECH XT Single-Sample Micro-Osmometer User Guide
4. Select the Allow deletion of users checkbox to
delete user information when a user is removed
from the active list.
5. Select the password strength options.
6. Press SAVE .
Configuring testing access
Authorized users can configure the following testing
requirements from the TESTING tab:
Require users to enter a user ID and/or a sample ID
before testing a sample.
Require login for users authorized to change the
operating range.
Allow all Administrators to test, calibrate, and change
the operating range.
NOTE: To configure user access to testing, calibration,
and changing the operating range, refer to the
Perform Calibration checkbox in Assigning user
restrictions.
1. Press RESTRICTIONS > TESTING.
2. Select the Enable Administrators access to perform
testing checkbox to authorize all Administrator-level
users to perform tests.
3. Select the User ID checkbox to require users to enter
a user ID before testing.
4. Select the Sample ID checkbox to require users to
enter a sample ID before testing.
NOTE: If the Review of Results and/or Approval of
Results checkbox is selected on the
Auditing tab, the Sample ID and User ID
checkboxes are selected automatically. To
disable the Sample ID and/or User ID
requirements, you must first disable Review
of Results and Approval of Results (see
Configuring audits of test results).
5. Select Auto Index Sample ID with Instrument Name
to enable auto indexing of sample IDs during testing.
When enabled, each sample is identified using the
instrument name and date with a sequential numeric
test index appended for each test run on that date.
In the example below:
The instrument name is QClab4A
The date is May 23, 2021
This is the second test (_2) run on this instrument
on this date
NOTE: If the instrument name has not been
changed since the instrument was first
installed, change the default name in order
to enable auto indexing (see Home screen).
6. Configure Change Range settings as needed:
Select Require Login to require selected users to
log in before changing the operating range.
Select the user level(s) required to log in before
changing the operating range.
Refer to Configuring the operating range for instructions.
7. Select the Enable Administrators access to perform
testing checkbox to
allow all Administrators to test
samples, calibrate the instrument, and change the
operating range.
8. Press SAVE.

Table of Contents

Other manuals for Advanced Instruments OsmoTECH XT

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Advanced Instruments OsmoTECH XT and is the answer not in the manual?

Advanced Instruments OsmoTECH XT Specifications

General IconGeneral
BrandAdvanced Instruments
ModelOsmoTECH XT
CategoryMedical Equipment
LanguageEnglish

Related product manuals