Chapter 3 Configuring settings
42
OsmoTECH XT Single-Sample Micro-Osmometer User Guide
4. Select the location of the exported data.
• To save the data to a USB drive, select USB and
insert the USB drive into the instrument’s USB
port.
Exported files are saved to the Advanced
Instruments>Export folder automatically created
on the USB drive.
• To save the data to a network folder:
a. Select Network.
b. Press FOLDER.
c. If a network folder has already been configured
for another data management function such as
database import or backup, press
to
select a folder configuration to copy to the
Export Folder screen.
Otherwise, enter the network folder settings
using the keyboards (
).
d. Press TEST to verify that the instrument can
access the folder.
NOTE: Contact your network administrator if the
instrument cannot access the folder over the
network.
e. Press SAVE and then EXIT.
5. Select the Require LOGIN for EXPORT checkbox to
require the selected users to login before exporting
data.
6. Press SAVE.
Exporting data manually
To export to the configured location (USB or network
folder) as a CSV file, complete the following steps.
1. Select the database table to export (Results, Users,
Events, or Controls.
2. Press EXPORT.
If exporting to a USB drive, exported files are saved
to the Advanced Instruments>Export folder
automatically created on the USB drive.
Repeat these steps for each database table you want to
export.
Backing up the database
1. Press the DATABASE BACKUP tab.
2. Select the location (USB or Network) for the backed
up data.