OsmoTECH XT Single-Sample Micro-Osmometer User Guide
Chapter 5: Calibration
IMPORTANT: Calibrate the OsmoTECH XT only with
Advanced Instruments Calibration Standards.
Using calibration standards that are not manufactured
by Advanced Instruments may yield inaccurate results
and is not recommended.
5.1 Factory calibration
Your OsmoTECH XT was factory-calibrated before it
shipped using five samples of each calibration standard
in the low range (0, 500, and 1500 mOsm/kg HO) and
in the high range (1500, 3000 and 4000 mOsm/kg HO).
The factory calibration is stored within its memory.
Therefore, you do not need to calibrate the instrument
upon installation.
5.2 When to recalibrate
Advanced Instruments recommends that you recalibrate
the OsmoTECH XT in any of the following cases:
If the test results for the reference solutions are out of
specification
NOTE: Advanced Instruments recommends that you
test appropriate reference solutions daily to
verify the calibration. For more information,
see Daily procedure.
If the instrument has been serviced (especially if any
instrument hardware was replaced)
If the ambient temperature has changed more than
5°C since the last calibration
In any other circumstances where calibration is
recommended or required based on the regulations
that govern your laboratory
5.3 Configuring the operating range
The OsmoTECH XT has a built-in calibration function
that by default guides you through testing five samples
each of three calibration standards.
Authorized users can select the operating range used to
calibrate the instrument from the Home screen.
NOTE: The selected range determines both the
calibration sequence and the operating range
used to test samples.
Select Low to calibrate the
instrument using 0, 500, and
1500mOsm/kg H
2
O calibration
standards.
Select High to calibrate the
instrument using 1500, 3000, and
4000 mOsm/kg H
2
O calibration
standards.
NOTE: Supervisors can either:
Allow all operators to calibrate the instrument, test
samples, and change the operating range, or
Restrict access to these features to supervisor-level only.
For more information, see Assigning user restrictions.

Table of Contents

Other manuals for Advanced Instruments OsmoTECH XT

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Advanced Instruments OsmoTECH XT and is the answer not in the manual?

Advanced Instruments OsmoTECH XT Specifications

General IconGeneral
BrandAdvanced Instruments
ModelOsmoTECH XT
CategoryMedical Equipment
LanguageEnglish

Related product manuals