4 Agilent 1100 Series LC/MSD Trap System Quick Start Guide
What is new in version 5.2 software?
New for overall software package
Data and methods
from previous
versions are
compatible with
5.2 software.
• Capacity to add on Security Pack 1.0 software for 21 CFR Part 11 compliance
• Supported on Windows 2000 SP4 and Windows XP SP1a only.
• Improved method structure for LC/MS systems including SOPs for LC, MS, and
processing parts of methods.
• SW Installation Qualification (IQ) report as text file.
New for individual programs with compliance software
ChemStation, Trap Control, DataAnalysis, QuantAnalysis
• Limited authorized access and authority checks
• User permissions and user roles
• Password policy
• Audit trails that track all operations for the system, method and sequence
setup, and analyses
• Audit trails that are time-stamped, secure, user-independent, readable,
automated and computer-generated
• Menu item ‘session locking’
• Menu item ‘User logon’ (to change user)
DataAnalysis, QuantAnalysis only
• Storage of multiple versions of electronic records (raw data, results, meta data,
etc.) with result revisioning
• Electronic signatures for results
• Reports with a listing of e-signatures and QuantAnalysis audit trail
New for individual programs with or without compliance software
LC/CE/Sequence control with ChemStation A.10.02
• Capacity to store methods in any location (recommended to use subdirectories
in the default method root)