Purification Solution - Developer's Guide 15
Preparing Default Purification Methods
2
General Method Settings
Flow Modulator (G7170B)
MSD
Access by Method > Edit entire method or by right click on MSD icon
Save the method as
General_Purification
Method > Save Method as
Splitting Operation Select Splitting disabled.
Set Up MSD Signals General
Peakwidth: 0.1 min (this defines the MSD data rate; make sure
that it is sufficient).
Active Signals: Mark 1 and 2 signals and name them, for
example,
Positive
and
Negative
MSD Signal Settings
Signal: 1, Mode: Scan, Polarity: Positive, Mass range: 125 – 725,
Step size: 0.1
Signal: 2, Mode: Scan, Polarity: Negative, Mass range: 125 –
725, Step size: 0.1
MSD type-specific settings MSD G6120B/G6125B
If the actual cycle time is above 0.9 s, consider decreasing it
by using only one scanning polarity (lower cycle time will
require less delay coil volume):
General > Active Signals > clear Signal 2
MSD Signal Settings > Signal 1 > specify polarity
MSD G6130B/G6135B
General > Peakwidth > 0.07 min

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity and is the answer not in the manual?

Agilent Technologies 1260 Infinity Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals