Purification Solution - Developer's Guide 85
Checkout Procedure
7
Acquire an analytical run using the Chemstation sequence
Acquire an analytical run using the Chemstation sequence
1 Set up a generic gradient in the analytical method:
a If using a combined analytical and preparative system:
a Use the Analytical_Purification method setting as described in
“Preparing Default Purification Methods” on page 9.
b Save the method as
Analytical_Gradient.
b Enter the recommended generic gradient in the method:
Pump > Solvents > B: 2 %
Pump > Timetable:
For an Agilent 1260 Infinity analytical run using a 4.6 × 50 mm, 5 μm
column and a Flow of 1 mL/min:
For an Agilent 1290 Infinity analytical run with a binary pump using a
2.1 × 50 mm, 1.8 μm column and a Flow of 0.6 mL/min:
NOTE
If the analytical run is to be acquired as a single run for the Purification software, it is
important to mark the Save Method with Data check-box in Method > Runtime Check
List. This applies to both the combined preparative instrument and the stand-alone
analytical instrument.
Time (min) A (%) B (%)
0.00 98 2
0.50 98 2
5.30 2 98
6.40 2 98
6.50 98 2
Stoptime: 9.0 min

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity and is the answer not in the manual?

Agilent Technologies 1260 Infinity Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals