24 Purification Solution - Developer's Guide
2
Preparing Default Purification Methods
Critical Method Parameters
Critical Method Parameters
In some situations, incorrect settings of the following parameters can cause 
incorrect functioning of the Purification software, incorrect fraction 
collection, or damage or contamination to some parts of the system. Therefore, 
if the purification system does not work correctly, ensure that the following 
parameters are set correctly:
Configuration • Fraction collector: Linked pump > select preparative pump
 General default
method
• Pump: Pressure limits (to avoid column damage)
• Autosampler: Needle wash (to avoid autosampler seat contamination)
• Autosampler: Vial/Well Bottom Sensing (use only with suitable sample wells 
or vials)
• Valve: Position > Use valve position (necessary for Purification software)
• Valve: Position switch at end of run > Do not switch (to avoid system 
overpressure due to flow path switch on the 2/10 valve)
• UV Detector: Signals > Set Signal A (set a wavelength suitable for analyzed 
compounds)
• MSD: Set Up MSD Signals > General > mark Active Signals (to collect MSD data)
• MSD: MSD Signal Settings > Mode > Scan (necessary for Purification 
software)
Analytical default
method
(combined
system)
• Valve 2/10 > Position > Use valve position > select correct analytical position
• Iso. Pump: Flow > 0 mL/min (make-up not used in analytical; flow can cause 
leak sensor error)
• UIB II: Expected ERI Mode > No mode check (for use with MSD, to avoid not 
ready state in some cases)
• Autosampler Plug: Plug solvent should contain only minimum amount of 
organic solvent to avoid unwanted partial elution of the sample on the 
column causing peak distortion and/or elution time variability