Purification Solution - Developer's Guide 19
Preparing Default Purification Methods
2
Preparative Method Settings
Preparative Method Settings
Load the general settings of the purification method:
General_Purification(“General Method Settings” on page 11)
Save the method as
Prep_Purification:
Method > Save Method As.
Isocratic Pump (G1310B)
Separated preparative MSD system — make-up pump
Quaternary Pump (G1311B)
Combined analytical and preparative MSD system — as make-up pump:
Prep Pump and Prep Pump Cluster (G7161A/B, G1361A and G1391A)
Flow 1.5 mL/min
Advanced Set Maximum Flow Gradient to about 10 times the target flow.
Channel D 100 %
Flow 1.5 mL/min
Advanced Set Maximum Flow Gradient to about 10 times the target flow.
Flow 20 mL/min (scale-up of 4.6 × 50 mm to 21.2 × 50 mm columns
with 5 µm particle size).
Advanced Channel A and B In Maximum Flow Gradient, set both Flow ramp up and Flow
ramp down to about 10 times the target preparative flow.

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity and is the answer not in the manual?

Agilent Technologies 1260 Infinity Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals