Purification Solution - Developer's Guide 11
Preparing Default Purification Methods
2
General Method Settings
General Method Settings
In all modules, set the Stoptime to As Injector/No Limit (infinite run time). The
actual stoptime will be specified when the method is used.
Isocratic Pump (G1310B)
Make-up pump for splitter used with separated preparative MSD instrument.
Quaternary Pump (G1311B)
Make-up pump for splitter and analytical pump for combined analytical and
preparative MSD instrument.
Prep Pump Cluster (G1361A and G1391A)
Stoptime Select As Injector/No limit
Pressure Limits Consider Active Splitter pressure limit (69 bar).
Stoptime Select As Injector/No limit
Pressure Limits Because of the 400-bar pressure limit of the preparative
system, set the upper limit for analytical scouting runs to 400
bar or lower.
Consider Active Splitter pressure limit (69 bar) when used as
make-up pump.
Solvent
description texts
Channel A: Water
Channel B: Acetonitrile
Solvent B 2%
Pressure Limits Set Max. limit. Find the pressure limit of the used columns (for
example, 21.2 mm guard system seals finger-tight up to 135 bar;
21.2 mm prep column without the guard up to 350 bar).
Stoptime Select As Injector/No limit

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity and is the answer not in the manual?

Agilent Technologies 1260 Infinity Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals