12 Purification Solution - Developer's Guide
2
Preparing Default Purification Methods
General Method Settings
Prep Pump (G7161A)
Prep Pump (G7161B)
Advanced Channel A Set Compressibility to 46 (Water)
Advanced Channel B Set Compressibility to 115 (Acetonitrile)
Solvents A
Select Aqueous and type the description text
Water
.
B
Mark the check box, enter
2 %
, select Acetonitrile, and type description text
Acetonitrile
.
Pressure Limits Set Max limit: find the pressure limit of the columns used (for
example, 21.2 mm guard system seals finger-tight up to 135 bar;
21.2 mm prep column without the guard up to 350 bar).
Advanced In Compressibility, keep the check box Use Solvent Types
marked
Solvents A
Select solvent line 1 or 2, select Aqueous and type the
description text
Water
.
B
Mark the check box, select solvent line 1 or 2, enter
2 %
, select Acetonitrile, and type description text
Acetonitrile
.
Pressure Limits Set Max limit: find the pressure limit of the columns used (for
example, 21.2 mm guard system seals finger-tight up to 135 bar;
21.2 mm prep column without the guard up to 350 bar).
Advanced In Compressibility, keep the check box Use Solvent Types
marked

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity and is the answer not in the manual?

Agilent Technologies 1260 Infinity Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals