Purification Solution - Developer's Guide 13
Preparing Default Purification Methods
2
General Method Settings
Dual Loop Autosampler (G2258A)
Prep Sampler (G7157A)
Valve 2/10 (G1170A and G4730A)
Selection of a flow path with a certain column.
Other Valves
Needle Wash Enable Needle Wash with Mode: Flush Port and Time15 s (to
avoid contamination of injection seat, do not use less than 10 s).
The wash procedure requires needle wash solution.
Stoptime Select As Pump/No limit
Advanced In Auxiliary, mark Vial/Well bottom sensing only if the vessels
used are suitable for this function, such as flat-bottomed glass
vials; see the Dual Loop Autosampler manual for more
information.
Needle wash Set Time15 s. The wash procedure requires needle wash
solution.
Stoptime Select As Pump/No Limit.
Position Select Use valve position
Position switch at end of run Select Do not switch
Position Names Enter description of preparative and analytical positions. For the
combined setup, set Position 1: Analytical, Position 2:
Preparative.
Position Select Use valve position (if applicable)