Purification Solution - Developer's Guide 13
Preparing Default Purification Methods
2
General Method Settings
Dual Loop Autosampler (G2258A)
Prep Sampler (G7157A)
Valve 2/10 (G1170A and G4730A)
Selection of a flow path with a certain column.
Other Valves
Needle Wash Enable Needle Wash with Mode: Flush Port and Time15 s (to
avoid contamination of injection seat, do not use less than 10 s).
The wash procedure requires needle wash solution.
Stoptime Select As Pump/No limit
Advanced In Auxiliary, mark Vial/Well bottom sensing only if the vessels
used are suitable for this function, such as flat-bottomed glass
vials; see the Dual Loop Autosampler manual for more
information.
Needle wash Set Time15 s. The wash procedure requires needle wash
solution.
Stoptime Select As Pump/No Limit.
Position Select Use valve position
Position switch at end of run Select Do not switch
Position Names Enter description of preparative and analytical positions. For the
combined setup, set Position 1: Analytical, Position 2:
Preparative.
Position Select Use valve position (if applicable)

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity and is the answer not in the manual?

Agilent Technologies 1260 Infinity Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals