EasyManua.ls Logo

Agilent Technologies 1260 Infinity - When to Use a Vacuum Degasser

Agilent Technologies 1260 Infinity
98 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
80 Purification Solution - Developer's Guide
6
Calibration Procedures
System Configuration and Delay Time Calibration
after the preparative sequence when the delay value needs to be written to
the collectors.
Insufficient UV-to-FC delay time or MSD-to-FC delay time
Insufficient UV-to-FC delay time
The minimum required delay time for UV detector triggered collection is:
4/frequency + Response time. The term 4/frequency stands for four data
points, which are needed in the slope-based collection.
There are several reasons why the UV-to-FC delay time may be too low:
The UV detector data rate is too low. A low UV data rate, especially in
combination with a high flow rate, may require a delay coil to ensure that
the UV detector has enough time to evaluate the FC trigger signal.
The volume of delay tubing in front of the fraction collector is too low.
The make-up flow is too low for the applied preparative flow. There are
limitations for combinations of preparative and make-up flows; see the
example in “System Configuration and Delay Time Calibration” on page 70.
The volume of tubing between the splitter and the detector(s) is too high.
There are several ways to increase the UV-to-FC delay time, but some of them
can have a certain disadvantage (given in parentheses), so consider the most
suitable option for the given situation:
Increase the UV data rate (a higher data rate typically has higher noise, so
in combination with the Slope mode for collection, the higher detector noise
can cause a false FC start/stop signal if the noise is visible along a collected
peak).
Use a longer delay coil in front of the fraction collector (a higher coil volume
causes more dispersion, so use coils with reduced dispersion rate such as
knitted coils; also, a longer coil with the same inner diameter will increase
back-pressure).
Decrease the preparative flow (influences chromatographic resolution).
Increase the make-up flow (increases the speed of the analytes when
passing through the detectors, which typically results in narrower and/or
smaller peaks, so consider whether or not the detector data rate is
sufficient for the signal response).

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity

Related product manuals